Model:FT19
Use:in-container sterilisation of food products and microbiological media.
Description
The autoclave is heated by an electrical element. The temperature inside the vessel is measured using a PT100 probe. The time and temperature
for the autoclave cycle are set by the operator.
There are four preprogrammed time temperature cycles:
- 30 minutes at 115°C
- 60 minutes at 115°C
- 15 minutes at 121°C
- 30 minutes at 121°C
There is also one variable cycle that can be set by the operator and is stored in memory. Operating temperature range is 100-121°C and cycle
time is 1-99 minutes.
Once the sterilisation cycle has been defined and the start button pressed, the unit automatically runs through the cycle. An audible beep
informs the operator when the cycle is complete and a second beep indicates a safe temperature has been reached to open the vessel. The
vessel is held closed by six locking points, which prevent the vessel from being opened when it is under pressure.
The FT19-10 option provides a stainless steel wire mesh basket with lifting handle (240mm Ø x 400mm deep).
Technical specifcations | |
Pressure adjustment | Adjustable safety valve |
Lid fastening | 6 locking points |
Safety | Resettable thermal cut-out |
Overall dimensions | |
Length | 0.39m |
Width | 0.37m |
Height | 0.49m |
Packed and created shipping specifcations | |
Volume | 0.1m3 |
Gross Weight | 24Kg |